Products

 

 

 

 

 

 

Products / Blackbodies / Ultra High Temperature Blackbodies / BB2000

Extended-Area Blackbody BB2000



   

Temperature range, K
Cavity diameter, mm
Cavity output opening diameter, mm
Cavity length, mm
Bottom apex angle
Effective emissivity
for 600 nm
Radiance spatial uniformity over 20 mm, %
Maximum current, A
Maximum voltage (DC), V
Service life at 2000 K, hrs
  800 - 2000
90

60
190
26°

0.999

1
200
40
200

The BB2000 wide-aperture blackbody was designed for calibrating infrared radiometers and pyrometers.


Figure above shows a schematic cross-section of the BB2000. Its main distinguishing feature is indirect resistive heating. The heater comprises 12 carbon-based fiber strips, which are arranged in 2 connected in series groups, and surrounding the radiating cavity. The BB2000 can be used without an output window and, when operating near the upper limit of the temperature range, requires pure argon flowing near the cavity aperture.

Blackbody source BB2000 is employed at secondary standard of radiation temperature at VNIIOFI (Russia).

Publications

V.I.Sapritsky, B.B.Khlevnoy, V.B.Khromchenko, B.E.Lisiansky, S.N.Mekhontsev, U.A.Melenevsky, S.P.Morozova, A.V.Prokhorov, L.N.Samoilov, V.I.Shapoval, K.A.Sudarev, M.F.Zelener. Precision blackbody sources for radiometric standards - Applied Optics, 1997, v. 36, pp. 5403-5408
Sapritsky V.I., Khlevnoy B.B., Khromchenko V.B., Lisiansky B.E., Mekhontsev S.N., Morozova S.P., Shapoval V.I., Samoilov L.N., Prokhorov A.V., Zelener M.F. Precision blackbody sources for thermometry and radiometry. – 6th International Symposium on Temperature and Thermal Measurement in Industry and Science (TEMPMEKO’96). Torino, Italy, 10-12 September, 1996, pp. 321-326