Delivery Set
Basic Set
Vacuum Tight Enclosure
2.5 kg Ga cavity, 0.9999999 purity, with Cu insert
Temperature Measurement and Monitoring Platinum Resistance
Sensors
Standard Mount for Blackbody
Optional Accessories
Refrigerated Circulator Bath Set
Vacuum Pump Set
The solid angle of a radiation from the cavity is limited
by the optical baffle with the aperture diaphragm with a diameter
of 8 mm. Ga cell consists of a Teflon cavity, which is inserted
into a Teflon case. The Ga cell is hermetically sealed by
O-ring, which is fixed by a nut.
The Ga cell is inserted into a closely fitting thick walled
copper block, which is used to provide a uniform temperature
field.
Copper tube is reeled-up and soldered around the heat exchanger.
Vacuum insulation of an internal volume of the radiating cavity
from the heat-exchanger internal volume permits to evacuate
these volumes independently.
References
1) Sapritsky V.I., Khlevnoy B.B., Khromchenko V.B., Lisiansky
B.E., Mekhontsev S.N., Morozova S.P, Shapoval V.I., Samoilov
L.N., Zelener M.F., Prokhorov A.V. Precision blackbody sources
for radiometry and radiation thermometry – TEMPBEIJING ’97.
Proc. of the Int. Conf. on the Temperature and Thermal Measurement.
Beijing, China, October 6-10, 1997, pp. 141-146
2) Sapritsky V.I., Khlevnoy B.B., Khromchenko V.B., Lisiansky
B.E., Mekhontsev S.N., Morozova S.P., Shapoval V.I., Samoilov
L.N., Prokhorov A.V., Zelener M.F. Precision blackbody sources
for thermometry and radiometry. – 6 th International Symposium
on Temperature and Thermal Measurement in Industry and Science
(TEMPMEKO’96). Torino, Italy, 10-12, September, 1996, pp.
321-326
|